Model: Polartronic H532
NSX: Schmidt Haensch - Đức

Phân cực kế họat động đến 5 bước sóng, với độ phân giải và độ chính xác cao. Ứng dụng trong nghiên cứu y tế, dược phẩm

The high-resolution (0.0001 °) polarimeter H532 with up to 5 wavelengths and the highest precision!

Polarimeters of the Polartronic H series are automatic, high-resolution measuring instruments with the following characteristics:

  • Accurate measurement of the optical activity of liquids according to ICUMSA, O.I.M.L. And Australian Standard K 157, European and American Pharmacopoeia
  • For the highest demands in quality and research laboratories
  • High precision over the entire measuring range by direct measurement of the angle
  • Short measuring times, no mechanical wear
  • Suitable for microtubes
  • Can be used for various applications with 5 wavelengths; Standard wavelengths are: 365nm, 435nm, 546nm, 579nm and 589nm (others on request)
  • Automatic temperature correction when using a corresponding tube
  • Calibration-free by spectral lamps
  • Safe and documented results through fully automatic and digital, objective and fast measurement
  • 10 methods and 10 scales programmable
  • Continuous measurement without START and RESET
  • Display of optical density
  • Determination of the optical rotatory dispersion (ORD), measurement at several wavelengths
  • RS232 to Ethernet adapter for data transfer to the PC
  • Extensive control software for test routines and configuration of the device
  • GLP / GMP compliant documentation
  • 21 CFR part 11 ready, corresponding software “Aquisys 2008” optionally available
  • Robust metal housing
  • High quality “Made in Germany” as well as advice and service on site possible
  • Internal measured value memory for 999 values

Technical data:
 

Measurement scales °Optical rotation, °Specific rotation, °Z International Sugar Scale, % Concentration (g/mL, g/100mL, g/L) 
7 scales freely
Measuring range ±89° (4 quadrants)
Resolution 0.0001° / 0.01°Z
Accuracy 0.002° / ± 0.01°Z
Reproducibility 0.001° / ± 0.01°Z
Sensitivity Up to OD 3
Wavelength Up to 5 wavelengths fixed: 365, 435, 546, 578, 589 (others on request)
Response time ≤ 4 sec. over the entire measuring range

Temperature

Temperature measurement If temperature sensor with appropriate tubes is used / T-Cell Peltier tubes
Temperature sample control with appropriate tubes via external water bath / Peltier temperature controlled sample tubes
Temperature control range depends on the water bath / 18 - 40 °C
Temperature resolution depends on the water bath / 0.01 °C
Temperature accuracy depends on the water bath / ± 0.03 °C
Temperature reproducibility depends on the water bath / 0.05 °C

Additional information

Light source Spectral lamps, calibration free
Display LCD display, monochrom
Operation Alpha numerical keybord, 16 x 16 characters
Operation TOUCH version Alpha numerical keyboard
Interface / Communication 2x RS232, 1x parallel, 1 PS2, USB (adapter)

Conformity

Conformity nternational Pharmacopoeia, OIML, ASTM, ICUMSA, Australian Standard K157