MODEL: AGERA
MAIN FEATURES
– 0º/45ºc Circumferential Geometry
- The only optical geometry that replicates how the human eye sees color
- The only dual-beam benchtop spectrophotometer that provides this geometry with circumferential viewing to eliminate sample directionality and deliver superior accuracy and consistency
– Full Spectrum, Balanced LED Illumination with UV Control: 360-700nm illumination
- Measure and quantify fluorescent and optically brightened samples
- Factory calibrated UV with user option to calibrate to their specific fluorescent standard
- Simultaneous UV-included and UV-excluded comparative data viewing and reporting
– Measured Area of View: The Agera port plate openings are oversized by approximately 3.174 mm (1/8 in) so that the illuminated and viewed areas of the sample are clear of the port edges. The measured area of the Agera is controlled by the lenses in the system and not controlled by the port plate opening size. This is an advantage over other instruments in that the port plate does affect the measurement.
-Sample Viewing and Image Capture
- Includes built-in 5-megapixel camera with 45-degree,full spectrum, balanced illumination
- Use the live on-screen image for precise sample placement
- Capture and store a true 0-degree field of view sample image as part of the data record
- Recalled images display an RGB histogram for image analysis
-Smart User Interface and Data Management
- View measurement results and color graphics on the 7-inch 1280×800 capacitive industrial touchscreen
- Easily communicate data where it needs to go via USB, USB on-the-go, Ethernet, and built-in email
– EasyMatch Essentials QC Software (21CFR PART 11)
- Easily navigate the embedded EasyMatch Essentials Quality Control Software which includes most major scales and indices used in industry
- Configure your specific workflow to streamline software navigation for your application
- Electronic Records option available (for FDA CFR21, Part 11 compliance)
– Superior Engineering
- Close Tolerance Spectrophotometer out of the box – no need for additional profiling software
- Protected optics secure your investment – designed to withstand the rigors of harsh environment
- Provides both color and 60-degree gloss results in a single measurement
- International gloss standard conformance to ASTM D523 and ISO2813 with Certificates of Traceability.
SPECIFICATION
Measurement Principle | Dual-beam Reflectance Spectrophotometer / Gloss meter |
Geometry | 0°/45°c (circumferential) ASTM E1164 |
Measurement Method | Port up or Port forward |
Read Time | < 3 sec |
Gloss Geometry | 60-degree (60º) |
Image Capture | High-resolution, D65 illuminated, 0°/45° image viewing, image capture and image recall |
Port Opening |
|
Area Measured |
|