DISSOLUTION MEDIA HEATING, DEGASSING, AND DISPENSING EZFILL+ – DISTEK
Model: DISTEK EZFILL+
Manufacturer: Distek, Inc.
Origin: USA
Features:
- Fast, reliable, and consistent results
- In-line heating eliminates time delays and wasted energy from heating bulk media tanks.
- High vacuum ensures effective deaeration and precise media volume delivery in approximately 90 seconds.*
- Color touch screen display
- Run stored methods and view/print reports, including dispensing parameters and user information.
- Automatic error-proof media settings via methods
- Volumes and temperatures can be set manually or using pre-stored methods.
- Track media parameters for data integrity compliance
- Track dispensing volumes, temperatures, operator details, and dispensing time via stored reports.
- Variable dispense volumes and temperature
- User-selectable or method-defined volumes ranging from 250 mL to 1000 mL with temperatures from ambient to 45°C.
- Surfactant capable
- Compatible with media containing up to 2% surfactants, minimizing foaming issues present in other systems.
- Compact design
- Easily fits on a bench or can be moved between baths using the optional mobile cart.
Specifications:
Parameter | Specification |
Dispensing volume | 250 mL to 1000 mL |
Dispensing accuracy | The greater of 1% of set volume or ±5 mL |
Heating capacity | +20°C degree increase from starting point |
Degassing | Up to a maximum of 3.0 ppm, not less than 5.0 ppm* |
Input filter | 25 mm in-line filter holder |
Temperature control | Cartridge heater |
Temperature accuracy | ±2°C |
Display | 5.7” color touch screen |
User management | Up to 50 users with multiple access levels |
Program modes | Manual and automatic (up to 100 pre-programmed methods) |
PC interface | USB, Ethernet |
Dimensions (ezfill+) | 31 cm (w) x 70 cm (h) x 26 cm (d) |
Dimensions (mobile cart) | 53 cm (w) x 94 cm (h) x 70 cm (d) |
Weight (ezfill+) | 19 kg |
Weight (mobile cart) | 15 kg |
Electrical power | 115V ± 15V 50/60Hz 15A or 230V ± 15V 50/60Hz 8A |
*Cycle time and degassing are temperature-dependent. Test conditions: water at 41°C.
Options