CATEGORY:
• Flame and graphite furnace integrated series
PRODUCT FEATURES:
• Flame and graphite furnace integrated machine; double atomizer series integrated design; no switching; users do not need to frequently adjust the position of graphite furnace injector;
• Dual light source design; flame and graphite furnace have independent light sources respectively, which overcomes the disadvantages of long optical path and large energy loss of traditional series all-in-one machine, and ensures that both analysis modes can have better optical design;
• Deuterium lamp background correction technology is adopted for flame and graphite furnace; specially imported deuterium lamp ensures its stable operation in modulation mode; strong energy and stable luminescence;
• With flame emission function, it can measure easily-ionized metal elements by emission method;
• Flame and graphite furnace atomization modes are automatically switched by software; relevant configuration and instrument conditions are automatically matched;
• Two sets of 8-Lamp slip ring rotating lighthouses can realize the simultaneous preheating of eight lights;
• It has the functions of power on self-test, automatic slit switching, automatic peak setting, automatic lamp position, automatic recommendation for analysis conditions, etc.
• Automatic energy balance technology, with D2 energy automatic attenuator, ensures that the cathode lamp is
matched within the scope of use, and ensures D2′ s service life and instrument’ s working condition;
• The graphite furnace has built-in power supply, longitudinal heating, light-controlled temperature rise and temperature calibration functions, and is designed with over temperature alarm of furnace body, low gas pressure alarm, broken or unclamped graphite tube alarm, etc;
• Equipped with 64 position graphite furnace automatic sampler; double syringe pump; vertical injection; visible injection process;
• The system of the flame part is stable and reliable with the design of integrated automatic water seal, automatic ignition, flame monitoring, gas leakage alarm, etc.;
• Equipped with gas shutdown button in case of accident;
• Can be equipped with hydride generator.
APPLICABLE FIELDS:
Food hygiene, geology, metallurgy, agricultural testing, chemical environmental protection, water quality testing, pharmaceutical commodity inspection, education and research, etc.
DETECTION ELEMENTS:
It is mainly used for more than 70 elements such as gold, silver, copper, lead, zinc, manganese, potassium, sodium, calcium, magnesium, iron and cadmium in food, surface water, groundwater, drinking water and other samples.
TECHNICAL SPECIFICATION:
|
Parameter |
Value |
| Wavelength Range | 190 – 900 nm |
| Monochromator | C-T (Czerny-turner)type; Focal length: 270mm |
| Optical Grating | 1800 pieces/mm |
| Spectral Bandwidth | 0.1, 0.2, 0.4, 0.7, 1.0, 1.6nm (automatic switching) |
| Wavelength Indication error | ≤ ±0.1 nm |
| Wavelength Repeatability | ≤ 0.1 nm |
| Spectral Bandwidth Deviation | ≤ ±0.02 nm |
| Baseline Stability(within 15 min) | Zero Drift Absorbance:≤±0.003Abs/15min; Absorbance of Instantaneous Noise: ≤0.003Abs/15min |
| Detection Limit | Flame: Cu ≤ 0.003 µg/mL; Graphite Furnace: Cd ≤ 0.5 µg/L* |
| Measurement Repeatability | Flame: Cu ≤ 0.6%; Graphite Furnace: Cd ≤ 2.0% |
| Linear Error | ≤ 10% |
| Background Deducting Capacity | ≥ 50 lần |
| Graphite Furnace Heating Range | room temperature~3000°C |
| Heating Rate | > 2000°C/s |
| Heating Mode | maximum power, step, rapid temperature rise |
| Temperature Control Accuracy | ≤ 5% |
| Temperature Stability | ≤ 3% |
| Sample Injection Mode | automatic injection |

