BIOLOGICAL SAFETY CABINET PURICUBE PRIME – CRYSTE

Origin: Korea

Brand:

Category: Tags: , Thương hiệu:

Biological Safety Cabinet (BSC) Class II with EN12469 and GS Mark certification by TÜV NORD refers to a highly specialized laboratory equipment unit that meets stringent European standards for safety and performance.

The BCS also featuring EASY – RUN mode by the ability to switch between normal and minimum airflow velocity by simply opening and closing the door, without the need to press any buttons during operation.

FEATURES:

Real Time – Smart Balancing Flow System: 

Real-time smart airflow velocity balancing is achieved through the dual positive pressure chamber (Patent Application No. 10-2517048) and dual sensor/blower system. This system protects both users and samples by ensuring the main airflow and inflow remain at a constant speed, while accurately and immediately detecting any changes in airflow velocity. Additionally, it automatically maintains the airflow velocity setting by supplementing the loading conditions of the HEPA filter according to the usage time, providing a safe environment identical to the initial state for a long time.

 

Ensuring Clean Air through Qualified Filters:

EN 1822-1 2019

Filter class
and group
Overall value
Efficiency (%) Penetration (%)
H14 (HEPA) ≥ 99.995 ≥ 0.005
U15 (ULPA) ≥ 99.9995 ≥ 0.0005

The standard HEPA filter of PURICUBE PRIME meets the EN1822 standard, which uses a highly reliable filter, accompanied by an individual report for leak testing.

Complete Response to Validation Requirements: PURICUBE PRIME has successfully passed the comprehensive EN12469 test, making it a trusted product that can seamlessly meet all field validation requirements.

 

SPECIFICATION:

MODEL PURICUBE PRIME 1200 PURICUBE PRIME 1500
ORDER CODE PCBP-12-MB PCBP-15-MB
Size Inner(W x D x H1)mm  1200 x 598 x 580   1500 x 598 x 580
Out(W x D x H2)mm  1324 x 798 x 2077   1624 x 798 x 2077
Base Stand(H)mm  670   670
Weight(kg) Main Top(Kg) 238 277
Bottom Stand(Kg) 34 38
Controller Display Back light blue LCD display
Function Downflow / Inflow Velocity Display, Light On/Off, UV lamp On/Off, Fan On/Off,
SMART OFF, Socket On/Off, Filter Life Span, Time/Timer, Easy Run Mode
Alarm Airflow Deviation / Unsafe Sash Height / Filter Life Exhausted /
UV lamp Exhausted / Power Failure & Secondary Evacuation Alarm
Material Worktop Stainless steel plate 304
Inside Stainless steel plate 304
Outside Epoxy powder coated steel plate
Sash Two-layer laminated toughened glass ≥ 6.38mm. Anti UV.
Tested Opening(H)mm 203 203
Max Opening(H)mm 480 480
Filterl H-14 HEPA filter : 99.995% at 0.3㎛ / Class 100 (ISO14644.1 class 5)
Type & Airflow system Class Ⅱ A2 (70% recirculation & 30% exhaust, Two-Flow Sensors)
Average Airflow Velocity Inflow  0.45±0.06 m/s  0.45±0.07 m/s
Downflow 0.30±0.06 m/s 0.37±0.07 m/s
Airflow Volume Inflow(m³/h) 420 540
Downflow(m³/h) 600 780
Exhaust(m³/h) 420 540
Air Valve 1EA
Socket 2EA with reinforced cover(mounted on inside wall)
UV Lamp 30W x 1EA
Illumination LED Light 15W x 2EA 20W x 2EA
Light Intensity ≥1000Lux
Noise 61.0dB(A) 62.1dB(A)
Values measured according to acoustic test in standard
EN 12469:2000 using standard ISO 11201. The uncertainty is K=2dB.
Power 220-240 V~, 50/60 Hz, 6.0A
Power Consumption 550W 700W
Fan Power 450W 600W

 