WATER PURE TYPE II + CLRW (CLSI) – OmniaLabDS – STAKPURE

Origin: Germany

Brand:

Category: Thương hiệu:

When safety is first priority and the quality of the purification decides the quality of results, then the OmniaLabDS system is the perfect solution. Even for large pure water quantities of up to 80 l/h, OmniaLabDS guarantees international water standards compliance. The combination of regenerative polishing cartridge and an optional emergency supply makes this system extremely reliable for supplying clinical analytical systems, as well as for feeding water to steam sterilizers and washer-disinfectors.

Model: OmniaLabDS

* Features:

  • OptiFill Dispenser is standard
  • 100L tank with quality recirculation and pressure outlet
  • Tank volume display in percent
  • Tank volume can be modularly increased
  • Simple, cost-effective filter replacement
  • Leakage sensor is standard
  • Emergency supply (optional)
  • Degassing unit (optional)

* Specifications:

OmniaLabDS 20

OmniaLabDS 40

OmniaLabDS 60

Pure water Typ II + CLRW (CLSI)

DIN EN 285 + ISO EN 15883

Pure water performance at 15°C [l/h]

20 40

60

Conductivity at 25 °C [µS/cm]

0.1 up to 1

Resistance at 25 °C [MΩ x cm]

10 up to 1

Dispensing performance dispenser [l/min.]

up to 2

Particles* > 0.2 µm [1/ml]

< 1

Bacteria* [KBE/ml]

< 0.1

Pure water tank pressurized outlet

optional

*with sterilizing filter 0.2 μm

* Feedwater Requirements:

OmniaLabDS 20/40/60

Softened or hardness-stabilized water according to DIN 2000
Feedwater temperature [°C]

+ 2 up to 35

Manganese and iron content [mg/l]

< 0.05

Free chlorine content [mg/l]

< 0.1

Silt density index (SDI)

max. 3

* Technical Data

OmniaLabDS 20/40/60

Connector size

R3/4″

Operating pressure [bar]

2 up to 6

Supply voltage [Volt/Hz]

90-240/50-60

Connected load [kW]

0.1

Ambient temperature [°C]

+ 2 up to 35

Dimensions tower* [W x H x D mm]

511 x 1520 x 575

Dimensions base cabinet tank [W x H x D mm]

511 x 800 x 575

Weight [kg]

39

* with OptiFill Dispenser

Sản phẩm đã xem

When safety is first priority and the quality of the purification decides the quality of results, then the OmniaLabDS system is the perfect solution. Even for large pure water quantities of up to 80 l/h, OmniaLabDS guarantees international water standards compliance. The combination of regenerative polishing cartridge and an optional emergency supply makes this system extremely reliable for supplying clinical analytical systems, as well as for feeding water to steam sterilizers and washer-disinfectors.