LyoMicron
Designed for laboratory / small-scale production.
Dimensions: 420 × 630 × 380 mm (-55°C) | 740 × 600 × 380 mm (-85°C).
Ice condenser capacity up to 6 kg.
Ice condenser temperature down to -85°C*.
Lightweight equipment, starting from 65 kg*.
Reliability and flexibility for every application.
- Ice condenser capacity up to 6 kg.
- Capacity up to 2,800 units of 2R vials.
Specifications
| Model | LyoMicron | LyoMicron |
|---|---|---|
| Ice condenser temperature | -55°C | -85°C |
| Dimensions (W × D × H) | 420 × 630 × 380 mm | 740 × 600 × 380 mm |
| Ice condenser capacity | 6 kg | 6 kg |
| Condensing capacity in 24 hours | 5 kg | 5 kg |
| Electrically heated shelves | Optional | Optional |
| Number of compressors | 1 | 2 |
| Power consumption | 1,500 W | 2,500 W |
| Weight | 65 kg | 85 kg |
| Vacuum control | Optional | Optional |
Base Software
By combining automation solutions with pharmaceutical freeze-drying processes, Dara-Lyo has developed a reliable, safe, and user-friendly system for monitoring and managing process data.
Within this product range, process data recording and recipe adjustment are particularly important. This enables users to identify, develop, and establish parameters aimed at future efficient production while achieving maximum performance for lyophilized products.
| Function | Description |
|---|---|
| Users | Allows registration of multiple user access levels through the HMI interface. |
| Recipe programming | Allows input of freeze-drying recipe parameters via the control screen. |
| Data export | Cycle data can be exported through a USB port. |
| Options | SCADA software for process data acquisition compliant with FDA 21 CFR Part 11 standards. |
Vacuum Pumps
Two-stage oil-sealed rotary vane vacuum pumps are used to achieve medium and low pressure. This equipment is highly suitable for R&D centers, laboratory environments, etc.
- Oil level sight glass.
- Maximum vacuum pressure: 2 × 10² mbar.
- KF 25 inlet connection.
- Flow rates from 6 m³/h to 36 m³/h.
- Durable and quiet operation.
- Easy maintenance.
- Hybrid pumps or dry pumps for organic solvents.
- Chemical resistance (optional).